Artigo Revisado por pares

Dosimetry on the spacelab missions IML1 and IML2, and D2 and on MIR

1996; Elsevier BV; Volume: 26; Issue: 6 Linguagem: Inglês

10.1016/s1350-4487(96)00101-1

ISSN

1879-0925

Autores

G. Reitz, R. Beaujean, C. Heilmann, Joachim Kopp, M. Leicher, K. Strauch,

Tópico(s)

Radioactivity and Radon Measurements

Resumo

Detector packages consisting of plastic nuclear track detectors, nuclear emissions, and thermoluminescence detectors were exposed inside BIORACK during the Spacelab missions IML1 and IML2, in different sections of the MIR space station, and inside the Spacelab module at rack front panels or stowage lockers and in the Spacelab tunnel during D2. In addition, during D2, each Payload Specialist (PS) has worn three permanent detector packages; one at the neck; one at the waist; and one at the ankle. Total dose measurements, particle fluence rate and LET spectra, number of nuclear disintegrations and neutron dose from this exposure are given in this report. The results are compared to theoretical calculations and to previous missions results. The dose equivalent (total radiation exposure) received by the PSs were calculated from the measurements and range from 190 to 770 μSv d−1. Finally, a cursory investigation of results from a particle telescope from two silicon detectors, first used in the last BIORACK mission on STS76, is reported.

Referência(s)
Altmetric
PlumX