Artigo Revisado por pares

Comparison of a New Reusable Gynecologic Laparoscopic Electric Morcellator with a Disposable Morcellator: a Preliminary Trial

2009; Elsevier BV; Volume: 16; Issue: 5 Linguagem: Inglês

10.1016/j.jmig.2009.05.010

ISSN

1553-4669

Autores

Maria Ángeles Martínez‐Zamora, Camil Castelo‐Branco, Juan Balasch, Francisco Carmona,

Tópico(s)

Gynecological conditions and treatments

Resumo

We compared the efficiency and safety of a newly developed reusable morcellator (Olympus VarioCarve) with that of a disposable morcellator (GyneCare Morcellex) for gynecologic laparoscopic surgery. In group 1 (n=15 patients; 11 myomectomies and 4 supracervical hysterectomies), the laparoscopic disposable morcellator was used, and in group 2 (n=14; 11 myomectomies and 3 supracervical hysterectomies), the reusable morcellator was used. There were no statistical differences in total tissue weight. Morcellating time was significantly shorter in group 2, and the rate of morcellation was significantly greater in group 2. Significantly fewer and longer pieces of tissue were removed with the reusable morcellator. No iatrogenic organ damage was observed.

Referência(s)