Dialyzer Performance in the Clinic: Comparison of Six Low‐Flux Membranes
1999; Wiley; Volume: 23; Issue: 9 Linguagem: Inglês
10.1046/j.1525-1594.1999.06297.x
ISSN1525-1594
AutoresPeter G. Kerr, Angela Lo, Man moo Chin, Robert C. Atkins,
Tópico(s)Acute Kidney Injury Research
ResumoAbstract: The aim of this study is to assess the clinical performance of 6 different low‐flux dialysis membranes under steady‐state conditions in terms of urea and phosphate clearances. Ten stable hemodialysis patients were examined. The following dialyzers were studied, all in 1.5‐ to 1.6‐m 2 format: cuprammonium, cellulose acetate, cellulose diacetate, hemophane, polysulfone (low‐flux), and polysynthane. The following parameters were examined: urea reduction ratio, phosphate reduction ratio, “instantaneous dialyzer clearance” for urea and phosphate, and total amount of urea and phosphate removed in the dialysate over a 1‐week (three dialyses) period. Although there were differences between the membranes, all produced results within a narrow range. There was no one membrane that produced superior clearances in all categories. The cellulose acetate membrane was the least satisfactory membrane. Phosphate clearances were at best one third that of urea clearances. When choosing a low‐flux dialysis membrane, urea and phosphate clearances are so similar amongst different membranes that other criteria are likely to have a greater influence on the choice of membrane.
Referência(s)