Dosimetric evaluation of a new OneDose MOSFET for Ir-192 energy
2006; IOP Publishing; Volume: 51; Issue: 5 Linguagem: Inglês
10.1088/0031-9155/51/5/015
ISSN1361-6560
AutoresRajesh Kinhikar, Pramod Kumar Sharma, C. Tambe, Deepak D. Deshpande,
Tópico(s)Advanced Optical Sensing Technologies
ResumoThe purpose of this study was to investigate dosimetry (reproducibility, energy correction, relative response with distance from source, linearity with threshold dose, rate of fading, temperature and angular dependence) of a newly designed OneDosetrade mark MOSFET patient dosimetry system for use in HDR brachytherapy with Ir-192 energy. All measurements were performed with a MicroSelectron HDR unit and OneDose MOSFET detectors. All dosimeters were normalized to 3 min post-irradiation to minimize fading effects. All dosimeters gave reproducible readings with mean deviation of 1.8% (SD 0.4) and 2.4% (SD 0.6) for 0 degrees and 180 degrees incidences, respectively. The mean energy correction factor was found to be 1.1 (range 1.06-1.12). Overall, there was 60% and 40% mean response of the MOSFET at 2 and 3 cm, respectively, from the source. MOSFET results showed good agreement with TLD and parallel plate ion chamber. Linear dose response with threshold voltage shift was observed with applied doses of 0.3 Gy-5 Gy with Ir-192 energy. Linearity (R2 = 1) was observed in the MOSFET signal with the applied dose range of 0.3 Gy-5 Gy with Ir-192 energy. Fading effects were less than 1% after 10 min and the MOSFET detectors stayed stable (within 5%) over a period of 1 month. The MOSFET response was found to be decreased by approximately 1.5% at 37 degrees C compared to 20 degrees C. The isotropic response of the MOSFET was found to be within +/-6%. A maximum deviation of 5.5% was obtained between 0 degrees and 180 degrees for both the axes and this should be considered in clinical applications. The small size, cable-less, instant readout, permanent storage of dose and ease of use make the MOSFET a novel dosimeter and beneficial to patients for skin dose measurements with HDRBT using an Ir-192 source compared to the labour demanding and time-consuming TLDs.
Referência(s)