Hydrogen flame chemiluminescence detector for sulfate in aqueous solutions
1969; Elsevier BV; Volume: 47; Issue: 2 Linguagem: Inglês
10.1016/s0003-2670(01)95673-9
ISSN1873-4324
AutoresWalter L. Crider, Naomi P. Barkley, Marta J. Knott, Robert W. Slater,
Tópico(s)Gas Sensing Nanomaterials and Sensors
ResumoThe instrument described is capable of detecting sulfate ion in aqueus solution at concentrations as low as 2μg sulfate per ml of solution. It is demonstrated to be of potential value for use as a specific sulfate analyzer in the hydrogen peroxide determination of sulfer dioxide. When used in conjunction with this analytical procedure, the detaction limit for sulfur dioxide collected in I-h samples is estimated to be 0.03 p.p.m. (v/v).Advantages over other methods presantly in use are discussed. L≈appareil décrit permet de déceler les sulfates en solution aqueuse à des concentrations descendant jusqu'à 2μg sulfate par ml de solution. Il convient également comme "analyseur" de sulfate spécifique lors du dosage de l'anhydride sulfureux au moyen de peroxyde d'hydrogène. Utilisé en relation avec ce procédé, la limite de détection pour des échantillons de SO2 recueilli en une heure est estimée à 0.03 p.p.m. (v/v). On examine les avantages sur d'autres méthodes actuellement en usage. Es wird ein Gerät beschrieben, das fähig ist, Sulfationen in wässriger Lösung bei Konzentrationen von weniger als 2 μg Sulfatml Lösung, nachzuweisen. Es eignet sich als spezifischer Sulfatanalysator bei der Wasserstoffperoxydmethode zur Schwefeldioxidbestimmung. In Verbindung mit diesem analytischen Verfahren ergibt sich eine Nachweisgrenze für Schwefeldioxid von 0.03 p.p.m. (v/v). Vorteile gegenüber anderen Methode werden diskutiert.
Referência(s)