HPLC with fluorescence detection of urinary phenol, cresols and xylenols using 4-(4,5-diphenyl-1H-imidazol-2-yl)benzoyl chloride as a fluorescence labeling reagent†
1998; Royal Society of Chemistry; Volume: 123; Issue: 11 Linguagem: Inglês
10.1039/a804582h
ISSN1364-5528
AutoresKen’ichiro Nakashima, Shinobu Kinoshita, Mitsuhiro Wada, Naotaka Kuroda, Willy R. G. Baeyens,
Tópico(s)Toxic Organic Pollutants Impact
ResumoA simple and sensitive HPLC method for the determination of phenolic compounds, i.e., phenol (Phe), cresols (Cres) and xylenols (Xyls), was developed. After a pre-column fluorescence derivatization of these compounds with 4-(4,5-diphenyl-1H-imidazol-2-yl)benzoyl chloride (DIB-Cl) at 60 degrees C for 30 min, 11 DIB derivatives were successfully separated within 50 min with an ODS column using CH3CN-H2O-CH3OH (25 + 22 + 53, v/v) as the eluent. The detection limits of DIB derivatives at a signal-to-noise ratio of 3 ranged from 0.15 to 1.09 microM (0.2-1.6 pmol per 20 microliters). The precision of the proposed method for both within- and between-day assays of free and total phenol related compounds was satisfactory (RSD < 9.5%). By the proposed method, Phe and p-Cre could be detected in normal urine samples, and the calculated concentrations of free Phe and p-Cre in unhydrolysed urine samples were 1.5 +/- 1.3 and 23.9 +/- 24.3 microM and those of total Phe and p-Cre in hydrolysed urine samples were 87.3 +/- 81.2 and 200.7 +/- 195.4 microM (n = 21), respectively.
Referência(s)