Measurement Accuracy of a Stand-Alone Oscillometric Central Blood Pressure Monitor: A Validation Report for Microlife WatchBP Office Central
2012; Oxford University Press; Volume: 26; Issue: 1 Linguagem: Inglês
10.1093/ajh/hps021
ISSN1941-7225
AutoresHao‐Min Cheng, Shih‐Hsien Sung, Yuan-Ta Shih, Shao‐Yuan Chuang, Wen‐Chung Yu, Chen-Huan Chen,
Tópico(s)Hemodynamic Monitoring and Therapy
ResumoThe superiority of prognostic value of blood pressure (BP) measured at central aorta (CBP) over conventional brachial BP measured by cuff-based BP monitors has reignited the development of new noninvasive techniques for estimating CBP. The present study validated the accuracy of CBP measured by a newly developed stand-alone CBP monitor. The CBP monitor provided readings of brachial systolic BP (SBP), brachial diastolic BP (DBP), central SBP, and central pulse pressure (PP). Brachial PP and central DBP were calculated from the relevant readings. The accuracy of the brachial and central SBP, PP, and DBP was validated against the simultaneously recorded invasively measured central aortic SBP, PP, and DBP, according to the invasive standard requirements for the noninvasive brachial BP monitors from the Association for the Advancement of Medical Instrumentation (AAMI) in 85 subjects (255 measurements; age range, 30–93 years). The mean differences of cuff BP with reference to the invasively measured central SBP, PP, and DBP were −2.6±9.0, −8.6±11.2, and 6.1±7.0mm Hg, respectively, with the former two being obviously underestimated at high CBP and overestimated at low CBP. In contrast, the corresponding differences for the central SBP, PP, and DBP measured by the CBP monitor were −0.6±5.5, −0.4±7.0, and −0.2±6.5mm Hg, respectively, without obvious systematic bias. The distribution of measurement errors for central SBP, PP, and DBP surpassed the AAMI criteria. Central SBP, PP, and DBP can be measured accurately by a stand-alone automatic BP monitor.
Referência(s)