Suitability of fetal scalp electrodes for monitoring the fetal electrocardiogram during labour
1990; IOP Publishing; Volume: 11; Issue: 4 Linguagem: Inglês
10.1088/0143-0815/11/4/002
ISSN0143-0815
AutoresJ. Westgate, Robert D. F. Keith, John Curnow, Emmanuel Ifeachor, Keith R. Greene,
Tópico(s)Advanced Battery Technologies Research
ResumoThe authors examined the suitability of five commonly available scalp electrodes for collecting this signal by examining their physical and electrical characteristics, together with a randomised clinical trial in which the ECG trace quality was assessed in 50 patients. The frequency response of Copeland electrodes was such that they attenuate the ECG signal more than the baseline noise. Difficulties were experienced in obtaining optimum attachment and the long, semi-rigid design increased movement artefact resulting in significantly poorer quality ECG signals. Whilst the Hewlett-Packard double spiral electrode had a near ideal frequency response, certain design features made it difficult to apply and remain secure so the clinical signals were of intermediate quality. The Corometrics and Cetro single spirals had the most stable attachment to the scalp and a near ideal frequency response, so produced significantly better signal quality in the clinical trial. Currently, single spiral electrodes are the most suitable for electrocardiographic data collection.
Referência(s)