Realization of the medium and high vacuum primary standard in CENAM, Mexico
2005; IOP Publishing; Volume: 42; Issue: 6 Linguagem: Inglês
10.1088/0026-1394/42/6/s01
ISSN1681-7575
AutoresJ C Torres-Guzman, L A Santander, Karl Jousten,
Tópico(s)Experimental and Theoretical Physics Studies
ResumoA medium and high vacuum primary standard, based on the static expansion method, has been set up at Centro Nacional de Metrolog?a (CENAM), Mexico. This system has four volumes and covers a measuring range of 1 ? 10?5?Pa to 1 ? 103?Pa of absolute pressure. As part of its realization, a characterization was performed, which included volume calibrations, several tests and a bilateral key comparison. To determine the expansion ratios, two methods were applied: the gravimetric method and the method with a linearized spinning rotor gauge. The outgassing ratios for the whole system were also determined. A comparison was performed with Physikalisch-Technische Bundesanstalt (comparison SIM-Euromet.M.P-BK3). By means of this comparison, a link has been achieved with the Euromet comparison (Euromet.M.P-K1.b). As a result, it is concluded that the value obtained at CENAM is equivalent to the Euromet reference value, and therefore the design, construction and operation of CENAM's SEE-1 vacuum primary standard were successful.