Comparison of Blood Glucose Testing Using Reagent Strips With and Without a Meter (Chemstrips bG and Dextrostix/Dextrometer)

1981; American Diabetes Association; Volume: 4; Issue: 3 Linguagem: Inglês

10.2337/diacare.4.3.417

ISSN

1935-5548

Autores

Paul Kubilis, Arlan L. Rosenbloom, Dennis C. Lezotte, Janet Silverstein, Pat Cimino, Eric D Rosenbloom, C Harvey,

Tópico(s)

Metabolism, Diabetes, and Cancer

Resumo

Eighteen counselors at a children's diabetes camp examined 360 blood specimens using the Dextrostix/Dextrometer (DX/DM) and Chemstrip bG (CS) systems. Sixty-five percent of the DX/DM and 60% of the CS observations were within 20% of plasma glucose results determined with the glucose analyzer (P = 0.19). CS errors were underestimates below 150 mg/dl and overestimates above that level. DX/DM reading errors were almost exclusively underestimates.