Revisão Acesso aberto Revisado por pares

Contemporary Pacemaker and Defibrillator Device Therapy

2007; Lippincott Williams & Wilkins; Volume: 115; Issue: 5 Linguagem: Inglês

10.1161/circulationaha.106.618587

ISSN

1524-4539

Autores

Mark H. Schoenfeld,

Tópico(s)

Cardiac Arrhythmias and Treatments

Resumo

HomeCirculationVol. 115, No. 5Contemporary Pacemaker and Defibrillator Device Therapy Free AccessReview ArticlePDF/EPUBAboutView PDFView EPUBSections ToolsAdd to favoritesDownload citationsTrack citationsPermissions ShareShare onFacebookTwitterLinked InMendeleyReddit Jump toFree AccessReview ArticlePDF/EPUBContemporary Pacemaker and Defibrillator Device TherapyChallenges Confronting the General Cardiologist Mark H. Schoenfeld, MD Mark H. SchoenfeldMark H. Schoenfeld From the Cardiac Electrophysiology and Pacemaker Laboratory, Hospital of Saint Raphael, Yale University School of Medicine, New Haven, Conn. Originally published6 Feb 2007https://doi.org/10.1161/CIRCULATIONAHA.106.618587Circulation. 2007;115:638–653Intracardiac pacemaker and defibrillator implantations were initially described a half century and a quarter century ago, respectively.1,2 Originally conceived as devices for the treatment of symptomatic bradycardia and sustained ventricular tachyarrhythmias, these therapies have subsequently been applied to the management of an increasing host of arrhythmia and other indications, including the treatment of heart failure.3 Although underuse of this technology has occasionally been reported,4 it has become abundantly clear that the number of patients deriving benefit from device therapy has risen exponentially; cardiac electrophysiology and pacing has emerged as a subspecialty within cardiology to address the growing needs of cardiac rhythm management.5 It is, nonetheless, incumbent on the general cardiologist to become proficient in certain aspects of device therapy.6 The goals of the present review are to highlight some of the fundamental areas of knowledge essential for follow-up, to provide an update on newer applications of device therapy, and to address some of the more common challenges and clinical scenarios encountered by the general cardiologist.Changing Indications for Device TherapyThe implantable cardioverter-defibrillator (ICD) was originally applied to patients surviving life-threatening ventricular tachyarrhythmias—so-called secondary prevention. Subsequently, primary prevention trials identified other patients who were at increased risk for sudden cardiac death but who had yet to manifest such risk,7 thereby expanding the indications for ICD therapy. The most recent iteration of guidelines for device implantation has identified additional conditions in which pacemaker and/or ICD therapy are associated with an enhancement of quality of life and/or survival.3,8 In particular, new ICD indications for primary prevention include prophylactic implantation in patients with ischemic cardiomyopathy and ejection fraction (EF) less than or equal to 30% (MADIT-II [Multicenter Automatic Defibrillator Implantation Trial II] criteria),9 and in patients with Brugada syndrome.10 The potential importance of ICD therapy for patients with nonischemic cardiomyopathy, EF less than 35%, and New York Heart Association class II or III congestive heart failure (CHF) symptomatology was demonstrated by the SCD-HeFT (Sudden Cardiac Death in Heart Failure) trial,11 a primary prevention application published after the 2002 guidelines.Potential new applications of cardiac pacing have been recognized, such as pacing in patients with neurocardiogenic syncope.3 Ironically, this application has been the subject of more recent debate: No benefit was found in the Second Vasovagal Pacemaker Study.12 Cardiac resynchronization therapy (CRT) has been identified as a class IIA therapy for patients with medically refractory New York Heart Association class III or IV CHF patients with cardiomyopathy, QRS interval of greater than or equal to 130 ms, and EF less than or equal to 35%.13,14 This latter application has allowed for a "marriage" of technologies in 1 device (CRT-ICD) that allows for treatment of CHF and protection against sudden cardiac death in selected patients.Indications may flourish and, almost as rapidly, languish. The time-honored use of dual-chamber pacing to achieve physiological pacing15 has come under new scrutiny. This concept, as originally proposed, referred to pacing techniques that were designed to allow for normal cardiovascular physiology, preserve atrioventricular (AV) synchrony, and provide heart rate increases (rate adaptiveness) as required by exercise or other physiological demands.16 Although this improved symptoms in many patients and reduced the incidence of atrial fibrillation (AF) compared with single-chamber pacing, no benefit was identified in certain patient subsets,17–19 as in the PASE (Pacemaker Selection in the Elderly)17 and UKPACE (United Kingdom Pacing and Cardiovascular Events)18 trials. Furthermore, potential adverse consequences of right ventricular apical pacing have led to the evaluation of alternative sites for pacing the heart,20 and our definition of physiological pacing continues to evolve.21 All of these points attest to the importance of periodically revisiting and updating implantation indications and guidelines as new data from randomized clinical trials become available.7,8,22Follow-Up of the Patient After Device ImplantationMuch has been written on device follow-up, and training requirements have been established for the selection, implantation, and follow-up of cardiac implantable electronic devices.23–26 Six-month intervals for ICD follow-up seem to be safe27; the same applies to pacemaker follow-up. These follow-up intervals may be extended as a result of innovations in remote monitoring28,29 and device automaticity. The latter may be defined as the automatic regulation of device function by programmed algorithms (see below) that are based on patient conditions and pacemaker/ICD system conditions, without the need for clinician input.30 Transtelephonic monitoring of pacemakers has been available for nearly 3 decades. The technology has shown that there is a relatively long trouble-free period in the midst of a device's life; most issues arise early after implantation (wound healing, threshold changes) or toward the device's end of life.30 These lessons have been applied increasingly to the management of ICDs; depending on the manufacturer, remote monitoring may be enabled through Internet-based systems or through radiofrequency transmission from a transmitter in the ICD via a phone device to a service center.Nonetheless, there is no substitute for direct patient contact, with an essential role for the physician to uncover device-related issues.24 In addition to periodic routine visits, more urgent examination may be required, as in the case of an ICD patient with increasingly frequent ICD discharges (as opposed to a single ICD shock, which could be addressed through remote monitoring). History taking is crucial to ensure that symptoms prompting the original implantation remain ameliorated and that device therapy has not resulted in the creation of new symptoms, as in the so-called pacemaker syndrome.31 This syndrome, most commonly caused by the loss of optimal AV synchrony in patients with single-chamber ventricular pacing, consists of cardiovascular and neurological symptoms that include neck/abdominal pulsations, palpitation, fatigue, dyspnea, and/or presyncope; the pathophysiology reflects diminished cardiac output and is associated with signs of systemic hypotension, AV valvular regurgitation, unpleasant cannon A waves in the neck, and pulmonary/hepatic congestion.24Examination of the wound is important early after implantation to address wound-healing as well as potential infection; the latter is particularly serious because it is rare for a patient to overcome device-related infections without extraction of the entire system.24 Pacing-induced complications such as diaphragmatic stimulation or myopectoral stimulation/inhibition also may be identified. Examination of chest radiographs is occasionally necessary to confirm lead integrity and position.32 The importance of 12-lead electrocardiography cannot be understated, not only for assessment of rhythm but also for confirmation of the appropriate QRS complex–paced morphology associated with various pacing lead locations (typically left bundle-branch block morphology and left-axis deviation with conventional right ventricular apical pacing). This examination is particularly important with the advent of alternative-site pacing, where less typical ECG patterns are observed.20,33Magnet application to a pacemaker generator is a useful adjunct to follow-up and converts the pacemaker to the asynchronous pacing mode (not inactivation).23 The resulting asynchronous rate is manufacturer specific and changes over the life of the battery as it approaches a preset elective replacement indicator rate. Thus, magnet application is useful in following device longevity and in confirming capture at the programmed output. It may be used to temporarily induce asynchronous pacing in cases of electromagnetic interference (EMI), such as electrocautery. With EMI, external signals may be detected by the device and misconstrued as intracardiac, resulting in inhibition of pacing output, unless sensing is eliminated by either magnet application or device reprogramming. In contrast, magnet application to ICD generators does not affect pacing, but it may temporarily or permanently suspend antitachycardia therapies.The importance of device interrogation and programmability has been appreciated since the inception of pacemaker therapy.34,35 Unfortunately, there is no universal pacemaker/ICD programmer to date; the manufacturer-specific nature of programmers constrains the nonelectrophysiologist who may not be familiar with or have access to the seemingly vast array of programmers required for follow-up. Ironically, allied professionals, often industry employed, may know more about a particular device and its associated programmer than a given physician; guidelines for such assisted follow-up have been established, and the physician must remain responsible for the device management and the patient's overall care.36Interrogation of a device provides a wealth of information, including but not limited to programmed parameters, such as the pacing rates and pacing mode (Table 1)37; real-time lead and battery impedance, as well as trends; event counters and histograms detailing the percentage of time sensed/paced, automatic mode-switch episodes, and atrial and/or ventricular arrhythmias; real-time and stored intracardiac electrograms; and marker channels that indicate what a device thinks it has seen or is seeing (Figure 1). Interrogation of programmed settings is essential in explaining phenomena that may be misinterpreted as device malfunction. For example, if a pacing rate is lower than the programmed lower tracking rate in a correctly functioning pacemaker, the phenomenon may be understood if interrogation also reveals that a hysteresis function has been programmed, allowing the patient's native rhythm to be maintained and to dip down to lower rates before pacing is triggered at a faster rate; this reduces current drain and preserves native rhythm. TABLE 1. Pacemaker Mode CodeAdapted with permission from Bernstein AD, Daubert JC, Fletcher RD, Hayes DL, Luderitz B, Reynolds DW, Schoenfeld MH, Sutton R. The revised NASPE/BPEG generic code for antibradycardia, adaptive-rate and multisite pacing. Pacing Clin Electrophysiol. 2002;25:260–264.37. Copyright 2002 Blackwell Publishing.Position IChamber paced: A indicates atrium; V, ventricle; D, both chambersPosition IIChamber sensed: O indicates none (asynchronous); A, atrium,; V, ventricle; D, both chambersPosition IIIResponse of pacer to sensed activity: O indicates none; T, triggered/tracked; I, inhibited; D, both triggers and inhibits output in a particular chamber, depending on the sensed eventPosition IVRate modulation: O indicates none; R, rate adaptive VVIVentricular demand pacing; inhibits ventricular pacing output if native activity is sensed above lower ventricular demand rate VOOAsynchronous ventricular pacing (no sensing capabilities) VVTTriggered ventricular pacing (perceives native ventricular activity and triggers ventricular pacing spike to coincide with sensed activity) VVIRRate-adaptive ventricular pacing AAI, AAIRAtrial demand pacing, atrial rate-adaptive pacing DDD, DDDRDual-chamber pacing, dual-chamber rate-adaptive pacing DDIDDD pacing without atrial tracking; AV paces at lower demand rate and inhibits atrial pacing output if atrial activity is sensed but does not trigger ventricular pacing in response to sensing atrial activity above the lower demand rate DDIRDDI Rate-adaptive pacing VDDAtrial synchronous pacing; senses/tracks atrial activity (to trigger ventricular pacing); senses in ventricle to inhibit if native ventricular activity is sensed; does not pace in atriumDownload figureDownload PowerPointFigure 1. T-wave sensing, resulting in spurious detection of tachycardia; both QRS and T waves are sensed, resulting in double counting by the device and thereby triggering detection of a pseudotachycardia by the device. FS indicates fibrillation sensed (faster interval that falls within VF zone); VS, ventricular sensed; and EGM, electrogram.The inability to perform telemetry could indicate that the wrong manufacturer's programmer was used, that the telemetry wand has not been positioned adequately over the generator (some models also require a magnet applied to the wand), or that the generator has reached anticipated (or premature) end of life.38 Elevated lead impedances may suggest a fracture in the conductor coil(s), a loose-set screw site at the lead insertion into the header block, or a problem at the tissue–electrode interface39 (Figure 2). Reduction of lead impedance may indicate a breach in lead insulation, often seen at the site of transvenous insertion—the so-called subclavian crush syndrome.40 Analysis of impedance trends may actually provide historical insight as to when lead disruption may have occurred (Figure 3). Reduction in battery voltage and a rise in battery impedance are characteristic of generators over time, providing a method by which to assess generator longevity. Download figureDownload PowerPointFigure 2. Detection of false signals on ventricular electrogram at nonphysiological intervals because of make-and-break connections in setting of loose-set screw, where lead interfaces with pacer generator, associated with elevated impedance. AS indicates atrial sensed; V EGM, ventricular electogram.Download figureDownload PowerPointFigure 3. Graph of impedance trends in defibrillator patient, showing intermittent elevations of impedance attributable to make-and-break connections at header block of a pacemaker generator.Device counters may allow for the quantification of events such as high atrial rates, high ventricular rates, and automatic mode-switch episodes, suggesting the interim occurrence of arrhythmias such as AF, supraventricular tachycardia (SVT), ventricular tachycardia (VT), or ventricular fibrillation (VF). Far-field sensing of ventricular depolarization, T waves, or noise by the atrial channel may result, however, in a false-positive incidence of high atrial rate or mode-switch episodes (Figure 4). Thus, pacemaker memory functions do not obviate the need for Holter or event monitoring in all patients with suspected arrhythmias. However, longer durations of individual episodes, on the order of many minutes to days, especially if confirmed by intracardiac electrograms, are more likely to correspond to true atrial dysrhythmias and an increased association with clinical events such as thromboemboli.41,42 Although device counters give a rough indication as to arrhythmia history, stored intracardiac electrograms have been increasingly used for validation of the true arrhythmia in pacemakers43 and ICDs44 (Figure 5). Download figureDownload PowerPointFigure 4. Noise on atrial channel in setting of reduced atrial lead impedance from subclavian crush syndrome in a pacemaker, falsely sensed as atrial fibrillation and causing automatic mode switching. The 3 channels are marker channel, atrial electrogram, and ventricular electrogram. A indicates atrial paced; V, ventricular paced; P, atrial sensed; and AMS, automatic mode switch to ventricular-paced mode.Download figureDownload PowerPointFigure 5. Ventriculoatrial dissociation during ventricular tachycardia in defibrillator patient; upper electrogram is atrial, lower is ventricular, with ventricular rate more than twice that of atrial rate; the last panel demonstrates delivery of a 29.9-J shock with restoration of sinus rhythm and 1:1 conduction.Automatic Algorithms for Detection and Treatment: A Brief SynopsisIt is beyond the scope of this review to detail all the device-specific algorithms available for sensing, discrimination, pacing, and defibrillation. Certain algorithms have, however, been increasingly used to automatically facilitate device function and follow-up37; these algorithms warrant some understanding on the part of the general cardiologist.Automatic algorithms that are intended to enhance sensitivity and optimize arrhythmia detection invariably encounter challenges regarding under- and oversensing.45 This is particularly important for ICDs where low-amplitude fibrillatory signals from the ventricle must be adequately sensed while avoiding oversensing in sinus rhythm. ICDs typically have algorithms to automatically adjust the gain or sensing threshold to sense reduced signals; it is, therefore, uncommon for current ICDs to miss VF. Conversely, oversensing may result in spurious discharge or inhibition of pacing output in ICD patients who are pacemaker dependent (Figure 6). Download figureDownload PowerPointFigure 6. Myopotential sensing with inhibition of ventricular pacing output in patient with pacer dependence and complete heart block in an cardioverter-defibrillator, with spurious interpretation of myopotentials as fibrillatory signals (VF) triggering cardioverter-defibrillator charge; uppermost channel is atrial electrogram, middle is ventricular rate-counting electrogram, lower is shock electrogram, and marker channels are below. Note inhibition of ventricular signals.Algorithms that allow for rhythm discrimination have become essential to ICD management. Inappropriate or spurious device discharges may be observed in up to 40% of patients, typically from misclassification of sinus tachycardia or from rapidly conducting supraventricular arrhythmias.46 For single-chamber devices, ventricular rate has served as the primary determinant of arrhythmia detection. Additional features such as sudden-onset criteria have aided discrimination of sinus tachycardia (gradual in onset) from VT or VF (typically sudden in onset) by comparing the interval at the onset of the tachycardia against an average of the previous intervals. This will not, however, distinguish VT from SVT. Rate-stability enhancements allow for measurement of beat-to-beat intervals, noting that monomorphic VT is typically regular, whereas AF and polymorphic VT or VF have wider variability when beat-to-beat variability is assessed. This feature proves to be particularly useful in discriminating AF from monomorphic VT. These enhancements reduce the likelihood of delivering spurious shocks for SVT without undersensing or delaying treatment for VT. Beyond this, certain algorithms have been developed to assess morphological characteristics of the electrogram during a patient's baseline rhythm. The resulting template may then be compared with the morphological characteristics obtained during a tachycardia, thereby aiding in the discrimination of supraventricular and ventricular rhythms.47 Although it is unpleasant for a patient to receive spurious shocks for misclassified SVT, it is far more hazardous for VT to be undetected; this has led manufacturers to incorporate a programmable time limit for tachycardia duration that, when exceeded, will result in the delivery of VT therapy.In dual-chamber ICDs, the presence of an atrial lead allows for comparison of ventricular and atrial rates during baseline rhythm and tachycardia, as well as the relative timing of atrial and ventricular activity. Ventricular rates exceeding atrial rates are typical for ventricular tachyarrhythmias with AV dissociation. Intracardiac electrograms are particularly useful in this regard (Figure 5). The Detect SVT trial reported a reduction of inappropriate SVT detection from 39.5% in the single-chamber detection arm compared with 30.9% in the dual-chamber arm.46 Arrhythmia detection by dual-chamber ICDs may be enhanced by combining various discriminators.48A recent multicenter trial, Comparison of Empiric to Physician-Tailored Programming of Implantable Cardioverter-Defibrillators (EMPIRIC), reported that standardized empirical ICD programming was as effective as physician-tailored programming as long as the empirical strategy included avoidance of detection of nonsustained tacyhcardia, avoidance of detection of SVTs as VT, empirical antitachycardia pacing for slow and fast VTs, and high-output first shocks.49 Nonetheless, misclassification of arrhythmic events is an ongoing concern, and much work remains in the refinement of detection algorithms, particularly in delineating rapid SVTs.50,51For the sake of completeness, it should be acknowledged that automatic sensing algorithms have been applied to a host of sensors unrelated to either atrial or ventricular depolarization, allowing for rate-adaptive pacing that is particularly well suited to patients exhibiting chronotropic incompetence.24,52 Thus, activity sensors, accelerometers, minute ventilation, and QT-interval sensing have been used alone or in combination to provide for more physiological rate modulation.Beyond sensing and detection, automatic algorithms have also been increasingly used in the treatment and prevention of arrhythmias. Pacing for bradycardia has been facilitated by automatic capture threshold assessment. The determination of ventricular capture threshold may be made at programmed intervals during the day or on a beat-by-beat basis with corresponding adjustments in output.53 Periodic and frequent assessments of thresholds are important from a safety standpoint because thresholds may change as a function of multiple factors, including changes in autonomic tone, electrolyte and metabolic abnormalities, and concomitant antiarrhythmic drug therapy. On the other hand, the ability to automatically down-regulate outputs when improved thresholds are detected extends device longevity.Automatic modulation of the AV interval may be accomplished in various ways in current pacing systems. It has long been appreciated that optimal hemodynamics are achieved with shorter AV delay intervals after atrial-sensed compared with atrial-paced events,54 and hemodynamics may be further enhanced by rate-adaptive shortening of the paced AV interval at faster heart rates.55Conventional right ventricular apical pacing may contribute to cardiac desynchronization and foster or exacerbate CHF in pacer patients.20,56,57 Thus, in the DAVID (Dual Chamber and VVI Implantable Defibrillator) trial,56 committed dual-chamber pacing was associated with increased death or hospitalization for CHF in ICD patients with depressed EF compared with patients programmed to an intentionally slow backup ventricular-paced mode. MOST (Mode Selection Trial) identified an increased rate of CHF hospitalization associated with ventricular-paced rhythm in pacemaker patients with sinus node dysfunction.57 What options, then, are available to minimize cardiac desynchronization induced by right ventricular apical pacing? Reprogramming to atrial pacing mode is one alternative, recognizing, however, that there is a finite, albeit small progression to AV block in patients with sick sinus syndrome, approximating 0.6% of patients per year.58 Programming a fixed long AV delay may minimize ventricular pacing, but it is not always feasible.59,60 In recognition of this, some manufacturers have incorporated conduction search algorithms that automatically search for intrinsic ventricular events that, if not sensed during an extension of the AV interval, will resume stimulation at the programmed AV delay61 (Figure 7). Download figureDownload PowerPointFigure 7. Minimal ventricular pacing mode. Upper panel: When atrial is paced at 60 bpm, intrinsic conduction occurs with first-degree block, inhibiting ventricular pacing. Lower panel: Atrial pacing at 80 bpm: at this faster atrial rate, atrioventricular node conducts with second-degree block, with progressive prolongation of time between atrial-paced complex and ventricular-sensed complex. When the device detects nonconduction to the ventricle, it triggers an atrioventricular-paced beat (AP/VP) (seventh ventricular complex). AP indicates atrial paced; VS, ventricular sensed; and VP, ventricular paced.In an effort to mimic naturally occurring circadian rhythms, some devices allow for programming of algorithms that will allow for slower pacing rates nocturnally.62 Another hysteresis function allows for pacing to kick in only when an abrupt rate drop is detected, applied with mixed results in the management of patients with neurocardiogenic syncope.12 Rate-smoothing programs are designed to prevent sudden changes in ventricular cycle length that may be responsible for either inducing pause-dependent arrhythmias or exacerbating symptoms of cardiac irregularity that are often encountered in AF.Automatic mode switching is, undoubtedly, one of the most useful and widely applied algorithms currently available; it provides for the detection of atrial tachyarrhythmias and conversion from dual-chamber to single-chamber ventricular pacing to preempt pacer-mediated tachycardia, with reversion to dual-chamber pacing once the sinus mechanism has reemerged63,64 (Figure 8). Oversensing, which results in false-positive automatic mode-switching responses, may be observed, most commonly from far-field sensing of the end of the QRS complex by the atrial channel, especially in the setting of noise and low right atrial implants63 (Figure 4). An analysis of automatic mode-switching episodes may provide the clinician with an estimate of the incidence of atrial dysrhythmias, prompting clinical decisions such as institution of anticoagulation, antiarrhythmic drug therapy, or even ablation; the above limitations of algorithm analysis regarding sensitivity and specificity must, however, be considered. Download figureDownload PowerPointFigure 8. Demonstration of patient with atrial flutter, with atrial electrograms at cycle length of 200 ms, initially tracking every other flutter wave with triggered ventricular-paced beat (2:1 lock-in) (A), then temporary inhibition of pacing by programmer showing flutter waves (B), followed by automatic mode switch to ventricular demand pacing (C); upper tracing is surface recording, lower is atrial electrogram. MS indicates mode switch; AS, atrial sensed; AR, atrial refractory; and VP, ventricular paced.Finally, some commentary is warranted regarding the potential utility of pacing algorithms for the prevention and/or treatment of AF. A variety of different methodologies have been developed to increase the atrial pacing rate when the native rhythm emerges and to periodically reduce the search rate for intrinsic atrial activity (dynamic atrial overdrive),65 to transiently use higher rate pacing after mode-switch episodes or spontaneous atrial ectopy, and to pace rapidly alone or in combination with shock therapy to terminate established arrhythmias.66 In general, evidence-based data in this area are sparse, and much work is needed to clarify whether there is any role for the use of permanent pacing in the prevention or treatment of AF, as recently noted by a joint working group of the American Heart Association and the Heart Rhythm Society.67Alternative Site Pacing and CRTAs previously noted, there has been growing evidence that cardiac desynchronization induced by conventional right ventricular apical pacing may have deleterious effects on cardiac function and patient symptomatology, corresponding with an increased risk of heart failure and death.20,56,57,60 Programming techniques have been used to minimize right ventricular pacing by lowering the baseline rate, changing to atrial-paced mode, or extending the AV delay.60,61 Invasive strategies to approach this problem have included upgrading from right ventricular apical pacing to biventricular pacing,68,69 alternative or selective site pacing from the His bundle,70 and primary CRT with left ventricular or biventricular pacing in patients with CHF, as in the MUSTIC (Multisite Stimulation in Cardiomyopathies),13 MIRACLE (Multicenter InSync Randomized Clinical Evaluation),14 COMPANION (Comparison of Medical Therapy, Pacing and Defibrillation in Heart Failure),71 and CARE-HF (Cardiac Resynchronization-Heart Failure) trials,72 or in non-CHF patients requiring pacing for standard indications (the HOBIPACE [Homburg Biventricular Pacing Evaluation] study73).The general cardiologist may be called on to interpret chest radiographs and electrocardiograms that may become increasingly complex with these alternative pacing modalities. Recently, electrocardiographic demonstration of a ratio of 1 or greater in the R/S wave amplitude in lead V1 has been shown to reliably detect left ventricular capture in patients with cardiac resynchronization devices33 (Figure 9). Changes in morphology may also be appreciated during assessment of pacing thresholds in biventricular systems (Figure 10). Download figureDownload PowerPointFigure 9. Twelve-lead electrocardiograms of patient with left bundle-branch block (A) and with biventricular pacing (B); note tall R wave in lead V1 during cardiac resynchronization therapy.Download figureDownload PowerPointFigure 10.

Referência(s)