Accuracy of Capillary Whole Blood International Normalized Ratio on the CoaguChek S, CoaguChek XS, and i-STAT 1 Point-of-Care Analyzers

2008; Oxford University Press; Volume: 130; Issue: 1 Linguagem: Inglês

10.1309/dek41w141y0krn8a

ISSN

1943-7722

Autores

Brad S. Karon, Robert D. McBane, Rajeev Chaudhry, Lisa K. Beyer, Paula J. Santrach,

Tópico(s)

Erythropoietin and Anemia Treatment

Resumo

We evaluated the accuracy of capillary whole blood international normalized ratio (INR) on the CoaguChek S (Roche Diagnostics, Indianapolis, IN), CoaguChek XS (Roche Diagnostics), and i-STAT 1 (i-STAT, East Windsor, NJ) point-of-care (POC) analyzers compared with venous plasma INRs determined by a reference laboratory method. Overall agreement between POC and laboratory plasma INR was very good, with median bias between capillary whole blood and laboratory plasma INRs varying from 0.0 to –0.2 INR units on all devices. More than 90% of results on the CoaguChek XS and i-STAT 1 and 88% of CoaguChek S results were within 0.4 INR units of the reference laboratory method. The CoaguChek XS and i-STAT 1 demonstrated greater accuracy than the CoaguChek S as measured by the number of results that differed by more than 0.5 INR units from the reference method. Median bias between CoaguChek S capillary whole blood and laboratory plasma INRs changed over time, demonstrating the need for ongoing quality assurance measures for POC INR programs.