Evaluation of Geometrical Sensitivity for Respiratory Motion Gating by GATE and NCAT Simulation

2007; Institute of Electrical and Electronics Engineers; Linguagem: Inglês

10.1109/iembs.2007.4353254

ISSN

1557-170X

Autores

Jianfeng He, Graeme O’Keefe, Gareth Jones, Timothy Saunder, Sylvia Gong, Moshi Geso, Andrew M. Scott,

Tópico(s)

Advanced X-ray and CT Imaging

Resumo

Respiratory motion artifacts can be a significant factor that limits the PET image quality. To improve image quality, surveillance systems have been developed to track the movements of the subject during scanning. Gating techniques utilizing the tracking information, are able to compensate for subject motion, thereby improving lesion detection. In this paper, we present a gating method that utilizes the geometric sensitivity gating (GSG) of a 3D-PET scanner system operating in list event acquisition mode. GSG method has several advantages over the existing methods, it only uses LOR events and is non-invasive, no additional hardware device is required and there is no additional patient preparation required. Using GATE (GEANT4 Application Tomographic Emission) and NCAT (NURBs(Non Uniform Rational B-Splines) Cardiac Torso) software packages with an Allegro PET configuration, realistic simulations of respiratory motion demonstrate that GSG can be used for respiratory gating.

Referência(s)
Altmetric
PlumX