Artigo Revisado por pares

The phantom portion of the American College of Radiology (ACR) Computed Tomography (CT) accreditation program: Practical tips, artifact examples, and pitfalls to avoid

2004; Wiley; Volume: 31; Issue: 9 Linguagem: Inglês

10.1118/1.1769632

ISSN

2473-4209

Autores

Cynthia H. McCollough, Michael R. Bruesewitz, Michael F. McNitt‐Gray, Krista Bush, Thomas Ruckdeschel, J. Thomas Payne, James A. Brink, Robert K. Zeman,

Tópico(s)

Radiology practices and education

Resumo

The ACR CT accreditation program, begun in 2002, requires the submission of approximately 20 images, several completed data sheets and printouts of three Excel worksheets. The procedure manual is very detailed, yet participants unfamiliar with the program or having minimal CT experience have needed to redo aspects of their submission, or in some cases do not receive accreditation, due to mistakes made by the physicist. This review of the phantom portion of the ACR CT accreditation program supplements the ACR provided instructions with additional photos of phantom setup, region-of-interest (ROI), and image placement on the film sheets, and examples of completed portions of actual (but anonymous) submissions. Common mistakes, as well as uncommon but interesting images, are shown and explanations are given as to what could have been done to avoid the problem. Additionally, a review of CT dose measurement techniques and calculations will enable the physicist to better assist sites where typical exam doses are above the ACR reference values.

Referência(s)
Altmetric
PlumX