Artigo Revisado por pares

Reduced vascular trauma after temporary occlusion with modified biemer and yasargil clips

1981; Wiley; Volume: 2; Issue: 3 Linguagem: Inglês

10.1002/micr.1920020307

ISSN

1098-2752

Autores

Manuel Dujovny, Guillermo Munoz, Debra L. Nelson, Roberto Langhi, Jack M. Fein, Nir Kossovsky Ba,

Tópico(s)

Aortic Disease and Treatment Approaches

Resumo

Abstract The Biemer and Yasargil microvascular clips that are currently available commercially can traumatize the vascular endothelium. In an effort to resolve this problem, both clip designs were structurally modified by changing the length, width, and surface of the clip blades in various ways. Also, the occlusion force of the clips was set at either 15 or 30 g. Studies on the carotid arteries of 140 rats showed a correlation between the structural and mechanical alterations of the clips and the degree of endothelial trauma produced by temporary occlusion for 1 hour. The most favorable results were obtained with a Biemer clip that had long, wide, smooth blades and an occluding force of 15 g.

Referência(s)