Artigo Revisado por pares

Contrast-to-Noise Ratio and Low-Contrast Object Resolution on Full- and Low-Dose MDCT: SAFIRE Versus Filtered Back Projection in a Low-Contrast Object Phantom and in the Liver

2012; American Roentgen Ray Society; Volume: 199; Issue: 1 Linguagem: Inglês

10.2214/ajr.11.7421

ISSN

1546-3141

Autores

Mark E. Baker, Frank Dong, Andrew N. Primak, Nancy A. Obuchowski, David M. Einstein, Namita Gandhi, Brian R. Herts, Andrei S. Purysko, Erick M. Remer, Neil Vachani,

Tópico(s)

Advanced X-ray and CT Imaging

Resumo

The purpose of this article is to evaluate the effect of sinogram-affirmed iterative reconstruction (SAFIRE) on contrast-to-noise ratio (CNR) compared with filtered back projection (FBP) and to determine whether SAFIRE improves low-contrast object detection or conspicuity in a low-contrast object phantom and in the liver on full- and low-dose examinations.A low-contrast object phantom was scanned at 100%, 70%, 50%, and 30% dose using a single-source made of a dual-source MDCT scanner, with the raw data reconstructed with SAFIRE and FBP. Unenhanced liver CT scans in 22 patients were performed using a dual-source MDCT. The raw data from both tubes (100% dose) were reconstructed using FBP, and data from one tube (50% dose) were reconstructed using both FBP and SAFIRE. CNR was measured in the phantom and in the liver. Noise, contrast, and CNR were compared using paired Student t tests. Six readers assessed sphere detection and conspicuity in the phantom and liver-inferior vena cava conspicuity in the patient data. The phantom and patient data were assessed using multiple-variable logistic regression.The phantom at 70% and 50% doses with SAFIRE had decreased noise and increased CNR compared with the 100% dose with FBP. In the liver, the mean CNR improvement at 50% dose with SAFIRE compared with FBP was 31.4% and 88% at 100% and 50% doses, respectively (p < 0.001). Sphere object detection and conspicuity improved with SAFIRE (p < 0.001). However, smaller spheres were obscured on both FBP and SAFIRE images at lower doses. Liver-vessel conspicuity improved with SAFIRE over 50%-dose FBP in 67.4% of cases (p < 0.001), and versus 100%-dose FBP, improved in 38.6% of cases (p = 0.085). As a predictor for detection, CNR alone had a discriminatory ability (c-index, 0.970) similar to that of the model that analyzed dose, lesion size, attenuation difference, and reconstruction technique (c-index, 0.978).Lower dose scans reconstructed with SAFIRE have a higher CNR. The ability of SAFIRE to improve low-contrast object detection and conspicuity depends on the radiation dose level. At low radiation doses, low-contrast objects are invisible, regardless of reconstruction technique.

Referência(s)