Artigo Revisado por pares

High perimeter impedance defibrillation electrodes reduce skin burns in transthoracic cardioversion

1998; Elsevier BV; Volume: 82; Issue: 9 Linguagem: Inglês

10.1016/s0002-9149(98)00571-2

ISSN

1879-1913

Autores

Lawrence A. Garcia, Luis A. Pagan‐Carlo, Mary Seabury Stone, Richard E. Kerber,

Tópico(s)

Cardiac pacing and defibrillation studies

Resumo

During cardioversion, skin burns result from current preferentially flowing at the electrode edge. We tested new electrodes with high perimeter impedance to yield more uniform current distribution; these electrodes reduced histopathologic skin injury.