Maximizing the dynamic range of the Humphrey Field Analyzer for blue‐on‐yellow perimetry
1993; Wiley; Volume: 13; Issue: 4 Linguagem: Inglês
10.1111/j.1475-1313.1993.tb00500.x
ISSN1475-1313
AutoresChris Hudson, John M. Wild, J.A. Archer-Hall,
Tópico(s)Glaucoma and retinal disorders
ResumoThe conditions employed for blue-on-yellow perimetry (i.e. a blue stimulus filter and a high luminance yellow background) are in conflict with those that allow a maximum dynamic range. A modification is described for increasing the dynamic range of the Humphrey Field Analyzer (HFA) for blue-on-yellow perimetry. The position of the HFA stimulus bulb was lowered by reducing the length of both terminals. A concave mirror with a 28 mm radius of curvature was positioned behind the lowered stimulus bulb to give a filament image partly above and partly overlapping the original filament profile within the confines of the size V stimulus patch diameter. The modification achieved a 60% increase in the light output on the default 0 dB value of 10,000 apostilbs and, therefore, a useable dynamic range for blue-on-yellow perimetry of 2.94 log units. The effective increase in voltage supply to the stimulus bulb permitted a filament life of 260 operating hours. Furthermore, the modification ensured the correct calibration of the instrument.