Comparison of Two Rapid Strip Tests Based on IGFBP-1 and PAMG-1 for the Detection of Amniotic Fluid
2008; Thieme Medical Publishers (Germany); Volume: 25; Issue: 4 Linguagem: Inglês
10.1055/s-2008-1066876
ISSN1098-8785
AutoresFrank K. Chen, Joachim W. Dudenhausen,
Tópico(s)Pregnancy and preeclampsia studies
ResumoOur objective was to compare two rapid strip tests for the detection of amniotic fluid, based on the detection of insulin-like growth factor-binding protein-1 (IGFBP-1) and of placental α-microglobulin-1 (PAMG-1). Samples of amniotic fluid were taken in 20 pregnant women between 31 3/7 and 41 2/7 gestational weeks at elective cesarean section before delivery of the newborn. These samples were diluted with 0.9% saline solution in a dilution series down to concentrations of 1:320. Immunoassay strip tests were then compared in their ability to detect remaining concentrations of amniotic fluid. In 5 cases, both test methods showed the same results. In all remaining 15 cases, the test based on PAMG-1 proved to be superior by detecting amniotic fluid at least at one descending concentration below the test based on IGFBP-1. Thus, the rapid strip test based on PAMG-1 seems to be a more sensitive bedside test compared with the test based on IGFBP-1 for the detection of amniotic fluid.