Reagentless amperometric immunosensor for human chorionic gonadotrophin based on direct electrochemistry of horseradish peroxidase
2004; Elsevier BV; Volume: 21; Issue: 2 Linguagem: Inglês
10.1016/j.bios.2004.10.023
ISSN1873-4235
AutoresJin Chen, Feng Yan, Zong Dai, Huangxian Ju,
Tópico(s)Analytical Chemistry and Sensors
ResumoA novel amperometric immunosensor for determination of human serum chorionic gonadotrophin (HCG) was constructed by immobilization of HCG with titania sol–gel on a glassy carbon electrode and the direct electrochemistry of horseradish peroxidase (HRP) labeled to HCG antibody (HRP-anti-HCG). The morphologies of the HCG membrane were characterized to be chemically clean, porous and homogeneous. HRP-anti-HCG was functionally conjugated with the immobilized HCG after incubation in phosphate buffer (PBS) containing HRP-anti-HCG. A direct electron transfer of HRP with a rate constant of 1.35 ± 0.40 s−1 was observed at the HRP-anti-HCG-HCG/titania sol–gel membrane modified electrode in 0.1 M PBS pH 7.0. With a competitive mechanism the differential pulse voltammetric peak current of the immobilized HRP decreased linearly with an increasing HCG concentration from 2.5 to 12.5 mIU/ml in the incubation solution. The HCG immunosensor showed a detection limit of 1.4 mIU/ml, a good accuracy and acceptable precision and reproducibility with an intra-assay CV of 4.7% at 5.0 mIU/ml and an inter-assay precision of 8.1% obtained at 10 mIU/ml. The biosensor displayed a good stability in a storage period of 30 days.
Referência(s)