Methods, Accuracy and Reliability of ITER Conductor Tests in SULTAN
2009; IEEE Council on Superconductivity; Volume: 19; Issue: 3 Linguagem: Inglês
10.1109/tasc.2009.2018508
ISSN1558-2515
AutoresP. Bruzzone, B. Stepanov, R. Wesche, Y. Ilyin, R. Herzog, Marco Calvi, M. Bagnasco, Francesca Cau,
Tópico(s)Magnetic confinement fusion research
ResumoIn the last decade, a large number of high current, force flow superconductors have been tested as short length samples in the SULTAN facility. The object of the test ranged over transient stability, thermal-hydraulic behavior, AC losses, joint resistance and proof-of-principle for innovative conductor design. Recently, with the ITER cable-in-conduit conductors (CICC), the basic DC transport properties have been the focus of the SULTAN test. The critical steps of the sample assembly and instrumentation are described, with emphasis on the application of the temperature sensors, verification of the signal treatment chain and calibration. The post-processing and the data reduction are focused on the assessment of the current sharing temperature, T cs : the conventional method of electrical field threshold detection by voltage taps is compared with the current sharing power detection by steady state gas-flow calorimetry. The longitudinal strain state of the conductors is discussed through the results of strain gauges applied on the jacket. Eventually, the value of a certified conductor test is highlighted in the frame of the quality control for the ITER magnets.
Referência(s)