New objective classification system for nuclear opacification
1997; Optica Publishing Group; Volume: 14; Issue: 6 Linguagem: Inglês
10.1364/josaa.14.001197
ISSN1520-8532
AutoresD. D. Duncan, Oodaye B. Shukla, S. K. West, Oliver D. Schein,
Tópico(s)Glaucoma and retinal disorders
ResumoWe have developed an autonomous objective classification scheme for degree of nuclear opacification. The algorithm was developed by using a series of color 35-mm slides acquired with a Topcon photo slit-lamp microscope and use of standard camera settings. The photographs were digitized, and first, and second-order gray-level statistics were extracted from within circular regions of the nucleus. Classifications of severity were performed by using these features as input to a neural network. Training versus classification performance was tested by using photographs of different eyes, and test/retest classification reproducibility was evaluated by using paired photographs of the same eyes. We demonstrate good performance of the classifier against subjective assessments rendered by the Wilmer grading system [Invest. Ophthalmol. Visual Sci.29, 73 (1988)] and markedly better test/retest reproducibility.