Artigo Revisado por pares

Antimony Film Microelectrode for Anodic Stripping Measurement of Cadmium(II), Lead(II) and Copper(II)

2010; Wiley; Volume: 22; Issue: 14 Linguagem: Inglês

10.1002/elan.200900583

ISSN

1521-4109

Autores

Marija Slavec, Samo B. Hočevar, Lucie Baldrianová, Eva Tesařová, Ivan Švancara, Božidar Ogorevc, Karel Vytřas,

Tópico(s)

Advanced Chemical Sensor Technologies

Resumo

Abstract The antimony film microelectrode (SbFME) is presented for measuring trace heavy metal ions in combination with anodic stripping voltammetry. The SbFME was tested in model solutions of 0.01 M HCl containing Cd(II) and Pb(II) in the presence of dissolved oxygen. The microsensor exhibited good linear behavior in the examined concentration range 20 μg L −1 to 100 μg L −1 and LoD of 1.9 and 3.1 μg L −1 for Cd(II) and Pb(II), respectively. The suitability of SbFME for measuring trace levels of Cu(II) was demonstrated using the standard reference material of natural water exploiting the sensor's specific characteristic of low re‐oxidation signal for antimony.

Referência(s)
Altmetric
PlumX