Comparison of Measurement of Pupil Sizes Among the Colvard Pupillometer, Procyon Pupillometer, and NIDEK OPD-Scan
2006; Slack Incorporated (United States); Volume: 22; Issue: 9 Linguagem: Inglês
10.3928/1081-597x-20061102-04
ISSN1938-2391
AutoresClaire McDonnell, Martina Rolincova, Jan A. Venter,
Tópico(s)Glaucoma and retinal disorders
ResumoTo compare pupil sizes measured with the Colvard pupillometer, Procyon pupillometer, and NIDEK OPD-Scan.Pupil diameter was measured in 90 consecutive eyes from 55 patients under mesopic and scotopic light conditions with all three instruments.The mean scotopic pupil diameter was 6.3+/-0.98 mm with the Colvard pupillometer and 6.45+/-1.01 mm with the Procyon pupillometer. The mean mesopic pupillometer was 5.58+/-1.01 mm with the Procyon pupillometer and 6.26-0.99 mm with theThe mesopic pupil diameter measured with the NIDEK OPD-Scan is more consistent with the scotopic pupil diameter measured with the Procyon and Colvard pupillometers than the mesopic pupil diameter measured with the Procyon pupillometer.
Referência(s)