Paper Electrochemical Device for Detection of DNA and Thrombin by Target-Induced Conformational Switching
2014; American Chemical Society; Volume: 86; Issue: 12 Linguagem: Inglês
10.1021/ac501438y
ISSN1520-6882
AutoresJosephine C. Cunningham, Nicholas Brenes, Richard M. Crooks,
Tópico(s)Advanced Biosensing Techniques and Applications
ResumoHere, we report a strategy for the design of an inexpensive paper analytical device (PAD) for quantitative detection of oligonucleotides and proteins. Detection is based on the principle of target-induced conformational switching of an aptamer linked to an electrochemical label. This simple and robust method is well matched to the equally simple and robust characteristics of the PAD platform. The demonstrated limits of detection for DNA and thrombin are 30 nM and 16 nM, respectively, and the device-to-device reproducibility is better than ±10%. The PAD has a shelf life of at least 4 weeks, involves little user intervention, and requires a sample volume of just 20 μL.
Referência(s)