Analytical and clinical performance of the new Fujirebio 25-OH vitamin D assay, a comparison with liquid chromatography-tandem mass spectrometry (LC-MS/MS) and three other automated assays

2015; De Gruyter; Volume: 54; Issue: 4 Linguagem: Inglês

10.1515/cclm-2015-0427

ISSN

1437-4331

Autores

Lanja Saleh, Daniel M. Mueller, Arnold von Eckardstein,

Tópico(s)

Healthcare cost, quality, practices

Resumo

Abstract We evaluated the analytical and clinical performance of the new Lumipulse Total 25 hydroxy vitamin D (25(OH)D) levels were measured in 100 selected serum samples from our routine analysis with Fujirebio 25(OH)D assay. The results were compared with those obtained with LC-MS/MS and three other automated 25(OH)D assays (Abbott, Beckman, and Roche). The accuracy of each assay tested was evaluated against a Labquality reference serum panel for 25(OH)D (Ref!25OHD; University of Ghent). Intra- and inter-day imprecision of the Fujirebio 25(OH)D assay was <5%. Fujirebio 25(OH)D assay showed the highest correlation among the assays tested with the LC-MS/MS method (R=0.986). The mean relative bias obtained was –15.6% (Fujirebio), –12.7% (Beckman), –2.1% (Abbott) and 9.7% (Roche) as compared to LC-MS/MS. Comparison with the Labquality certified reference serum panel yielded a mean bias of –11.8% (Fujirebio), –14.1% (Beckman), 4.4% (Abbott) and 3.2% (Roche), respectively. Compared to LC-MS/MS, the sensitivity of different methods in detecting vitamin D insufficiency (<50 nmol/L) varied from 100% for the Fujirebio assay to 72.7% for Roche, and specificity ranged from 94.4% for Roche to 87.6% for Beckman. The Lumipulse