Artigo Revisado por pares

High-Performance Liquid Chromatography-Tandem Mass Spectrometry as a Reference for Analysis of Tacrolimus to Assess Two Immunoassays in Patients With Liver and Renal Transplants

1997; Lippincott Williams & Wilkins; Volume: 19; Issue: 6 Linguagem: Inglês

10.1097/00007691-199712000-00015

ISSN

1536-3694

Autores

Paul Salm, Paul J. Taylor, A. K. Clark, Glenda A. Balderson, Anthony Grygotis, Ross Norris, Stephen V. Lynch, Leslie M. Shaw, Susan M. Pond,

Tópico(s)

Neurological Complications and Syndromes

Resumo

The accuracy and imprecision of three assays used for therapeutic monitoring of tacrolimus were tested using blood-containing weighed-in amounts of the drug, an enzyme-linked immunosorbent assay (ELISA), a microparticle enzyme immunoassay (MEIA I), and a high-performance liquid chromatography-tandem mass spectrometry (HPLC-MS2) assay. Accuracy was acceptable for the HPLC-MS2 assay at all concentrations tested (<10% deviation) and for the ELISA at 1.0 and 4.0 µg/l. Accuracy was not acceptable for the ELISA at 15.0 and 50.0 µg/l or for the MEIA I at all concentrations tested. Imprecision was acceptable for the HPLC-MS2 assay at all concentrations tested (coefficient of variation < 10%), for the ELISA at 15.0 and 50.0 µg/l, and for the MEIA I at 15.0 and 50.0 µg/l. Imprecision was not acceptable for the ELISA at 1.0 and 4.0 µg/l or for the MEIA I at 1.0 and 4.0 µg/l. This assessment with weighed-in amounts of tacrolimus verified the HPLC-MS2 assay as a reference method. The performance of the two immunoassays with HPLC-MS2 was then compared in the clinical setting using blood from patients with liver (n = 30) and renal (n = 37) transplants. In the liver transplant group (127 samples), the range of tacrolimus concentrations measured by HPLC-MS2, ELISA, and MEIA I was 1.9 to 31.8, 2.1 to 35.0, and less than 0.1 to 36.5 mg/l, respectively. In the renal transplant group (129 samples), the ranges were 1.7 to 26.1, 1.9 to 24.4, and 0.9 to 28.5 µg/l, respectively. Compared with the HPLC-MS2, the ELISA had minimal bias (0.1 to 0.2 µg/l) but unacceptable variability in values (SD > 13%). The MEIA I had unacceptable bias (1.7-1.8 µg/l) and variability (SD > 23%). These data indicated that neither the ELISA nor MEIA I is interchangeable with HPLC-MS2. Moreover, in view of the current trend to reduce the therapeutic dose of tacrolimus, quantitative results using the MEIA I would not be obtainable during therapeutic drug monitoring in some patients in whom effective therapeutic concentrations can be less than 5.0 µg/l.

Referência(s)