Artigo Acesso aberto Revisado por pares

Fluorometric liquid chromatographic determination of aliphatic aldehydes arising from lipid peroxides.

1986; Pharmaceutical Society of Japan; Volume: 34; Issue: 4 Linguagem: Inglês

10.1248/cpb.34.1694

ISSN

1347-5223

Autores

Kyoji Yoshino, Toshiki Matsuura, Mitsuaki Sano, Shinichi Saito, Isao Tomita,

Tópico(s)

Analytical Chemistry and Chromatography

Resumo

A fluorometric method for the determination of aliphatic aldehydes using 1, 3-cyclohexanedione reagent and high-performance liquid chromatography (HPLC) was developed.Straight-chain aldehydes of C3-C10 were reacted with 1, 3-cyclohexanedione at 60°C for 1 h and the fluorescent decahydroacridine derivatives2, 3) formed were determined by HPLC on an ODS column. The calibration curves for the aldehydes were linear in the range of 0-10 ng and the relative standard deviation at the 5 ng level was about 2.3%.This method was sensitive enough to measure the amount of aliphatic aldehydes, including 4-hydroxynonenal, generated by microsomal peroxidation in rat liver.

Referência(s)