A 3D Kinematic Measurement of Knee Prosthesis Using X-ray Projection Images (Countermeasure to the Overlap between Tibio-femoral Component Silhouettes)
2006; Japan Society Mechanical Engineers; Volume: 72; Issue: 713 Linguagem: Inglês
10.1299/kikaic.72.168
ISSN0387-5024
AutoresShunji HIROKAWA, Shogo ARIYOSHI, Ayaka Sato, Teppei KAWATA,
Tópico(s)Advanced X-ray and CT Imaging
ResumoWe have developed a technique for estimating 3D motion of knee prosthesis from its 2D perspective projections. As Fourier descriptors were used for compact representation of library templates and contours extracted from the prosthetic X-ray images, the entire silhouette contour of each prosthetic component was required. This caused such a problem as our algorithm did not function when the silhouettes of tibio and femoral components overlapped with each other. Here we planned a novel method to overcome it ; which was processed in two steps. First, the missing part of silhouette contour due to overlap was interpolated using a free-formed curvature such as Bezier. Then the first step position/orientation estimation was performed. In the next step, a clipping window was set in the projective coordinate so as to separate the overlapped silhouette drawn using the first step estimates. After that the localized library whose templates were clipped in shape was prepared and the second step estimation was performed. Computer model simulation demonstrated sufficient accuracies of position/orientation estimation even for overlapped silhouettes ; equivalent to those without overlap.
Referência(s)