Assessing a novel polymer-wick based electrode for EEG neurophysiological research
2016; Elsevier BV; Volume: 267; Linguagem: Inglês
10.1016/j.jneumeth.2016.04.009
ISSN1872-678X
AutoresRita Pasion, Tiago O. Paiva, Paulo Pedrosa, Hugo Gaspar, Beatriz Vasconcelos, Ana Martins, Maria Helena Amaral, J. M. Nóbrega, Ricardo N.M.J. Páscoa, Carlos Fonseca, Fernando Barbosa,
Tópico(s)Neural dynamics and brain function
ResumoThe EEG technique has decades of valid applications in clinical and experimental neurophysiology. EEG equipment and data analysis methods have been characterized by remarkable developments, but the skin-to-electrode signal transfer remains a challenge for EEG recording.A novel quasi-dry system - the polymer wick-based electrode - was developed to overcome the limitations of conventional dry and wet silver/silver-chloride (Ag/AgCl) electrodes for EEG recording.Nine participants completed an auditory oddball protocol with simultaneous EEG acquisition using both the conventional Ag/AgCl and the wick electrodes. Wick system successfully recorded the expected P300 modulation.Standard ERP analysis, residual random noise analysis, and single-trial analysis of the P300 wave were performed in order to compare signal acquired by both electrodes. It was found that the novel wick electrode performed similarly to the conventional Ag/AgCl electrodes.The developed wick electrode appears to be a reliable alternative for EEG research, representing a promising halfway alternative between wet and dry electrodes.
Referência(s)