Artigo Acesso aberto Revisado por pares

Three-dimensional paper-based slip device for one-step point-of-care testing

2016; Nature Portfolio; Volume: 6; Issue: 1 Linguagem: Inglês

10.1038/srep25710

ISSN

2045-2322

Autores

Kwi Nam Han, Jong‐Soon Choi, Joseph Sang‐Il Kwon,

Tópico(s)

SARS-CoV-2 detection and testing

Resumo

Abstract In this study, we developed a new type of paper-based analytical device (PAD), the three-dimensional (3D) slip-PAD, to detect infectious human norovirus for global healthcare. The 3D configuration of the papers combined with a slip design provides unique features and versatility that overcome the limitations of fluidic manipulation and sensitivity in point-of-care (POC) tests. The assay can be carried out in a single step based on a moveable slip design, making it suitable for unskilled users. The 3D fluidic network developed by layered construction of wax-patterned papers provides different fluidic paths for the sequential delivery of multiple fluids without the need for peripheral equipment. The release and mixing of enhancement reagents on the device improved the sensitivity and detection limit. The assay results could be visualized by naked eye within 10 min, with subsequent amplification of the signal over time (<60 min). The device showed a broad dynamic range of detection and high sensitivity, with a detection limit of 9.5 × 10 4 copies ml −1 for human norovirus. These results demonstrate that the 3D slip-PAD is a sensitive diagnostic assay for detecting human norovirus infection that is particularly suitable for POC testing in regions where resources are scarce.

Referência(s)