Comparison of tissue oximeters on a liquid phantom with adjustable optical properties
2016; Optica Publishing Group; Volume: 7; Issue: 8 Linguagem: Inglês
10.1364/boe.7.002973
ISSN2156-7085
AutoresStefan Kleiser, Nassim Nasseri, Björn F. Andresen, Gorm Greisen, Martin Wolf,
Tópico(s)Neonatal Respiratory Health Research
ResumoThe SafeBoosC trial showed that cerebral oximetry combined with a treatment guideline can reduce the the burden of hypoxia in neonates by 50% [Brit. Med. J.350, g7635 (2015)]. However, guidelines based on oximetry by one oximeter are not directly usable by other oximeters. We made a blood-lipid phantom simulating the neonatal head to determine the relation between oxygenation values obtained by different oximeters. We calculated coefficients for easy conversion from one oximeter to the other. We additionally determined the corresponding SafeBoosC intervention thresholds at which we measured an uncertainty of up to 9.2% when varying hemoglobin content from 25μM to 70μM. In conclusion, this paper makes the comparison of absolute values obtained by different oximeters possible.