Artigo Revisado por pares

Experimental results with TOFPET2 ASIC for time-of-flight applications

2017; Elsevier BV; Volume: 912; Linguagem: Inglês

10.1016/j.nima.2017.11.034

ISSN

1872-9576

Autores

R. Bugalho, A. Di Francesco, L. Ferramacho, Carlos Leong, T. Niknejad, Luís B. Oliveira, L. Pacher, M. Rolo, Angelo Rivetti, Miguel Silveira, José Silva, Rui Silva, S. Tavernier, J. Varela,

Tópico(s)

Particle Detector Development and Performance

Resumo

We present the experimental results obtained with TOFPET2, a readout and digitization ASIC for radiation detectors using Silicon Photomultipliers. The circuit is designed in CMOS 110 nm technology, has 64 independent channels and is optimized for time-of-flight measurement in PET or other applications. The chip has quad-buffered TDCs and charge integration ADCs in each channel. The chip tape-out was done in September 2016 and first tests started in beginning March 2017. Coincidence Time Resolution (CTR) of 164ps FWHM has been measured with 22Na point source. The energy resolution achieved for the 511keV peak is 10.5% FWHM.

Referência(s)