Artigo Acesso aberto Revisado por pares

Phantom Study of In-Stent Restenosis at High-Spatial-Resolution CT

2018; Radiological Society of North America; Volume: 289; Issue: 1 Linguagem: Inglês

10.1148/radiol.2018180188

ISSN

1527-1315

Autores

Hiromitsu Onishi, Masatoshi Hori, Takashi Ota, Atsushi Nakamoto, Keigo Osuga, Mitsuaki Tatsumi, Hideyuki Fukui, Shinsuke Tsukagoshi, Ayumi Uranishi, Yasuo Saito, Akira Taniguchi, Yukihiro Enchi, Kazuhiko Sato, Noriyuki Tomiyama,

Tópico(s)

Renal and Vascular Pathologies

Resumo

Purpose To examine the diagnostic performance of high-spatial-resolution (HSR) CT with 0.25-mm section thickness for evaluating renal artery in-stent restenosis. Materials and Methods A 0.05-mm wire phantom and vessel phantoms with renal stents with in-stent stenotic sections of varying diameters were scanned with both an HSR CT scanner equipped with 160-section multi-detector rows (0.25-mm section thickness) and a conventional CT scanner. The wire phantom was used to analyze modulation transfer function (MTF). With the vessel phantoms, the error rates were calculated as the absolute difference between the measured diameters and true diameters divided by the true diameters at the narrowing sections. For qualitative evaluation, overall image quality and diagnostic accuracy for evaluating stenosis in three stages were assessed by two radiologists. Statistical analyses included the paired t test, Wilcoxon signed-rank test, and McNemar test. Results HSR CT achieved 24.3 line pairs per centimeter ± 0.5 (standard deviation) and 29.1 line pairs per centimeter ± 0.4 at 10% and 2% MTF, respectively; and conventional CT was 12.5 line pairs per centimeter ± 0.1 and 14.3 line pairs per centimeter ± 0.1 at 10% and 2% MTF, respectively. The mean error rate of the measured diameter at HSR CT (8.0% ± 5.8) was significantly lower than that at at conventional CT (16.9% ± 9.3; P < .001). Image quality at HSR CT was significantly better than that at conventional CT (P < .001), but HSR CT was not significantly superior to conventional CT in terms of diagnostic accuracy. Conclusion Compared with conventional CT, high-spatial-resolution CT achieved spatial resolutions of up to 29 line pairs per centimeter at 2% modulation transfer function and yielded improved measurement accuracy for the evaluation of in-stent restenosis in a phantom study of renal artery stents. Published under a CC BY 4.0 license.

Referência(s)
Altmetric
PlumX