Development of TiAu TES x-ray calorimeters for the X-IFU on ATHENA space observatory
2018; Linguagem: Inglês
10.1117/12.2313439
AutoresP. Khosropanah, E. Taralli, L. Gottardi, C. P. de Vries, K. Nagayoshi, M. Ridder, H. Akamatsu, M. P. Bruijn, J. R. Gao,
Tópico(s)Superconducting Materials and Applications
ResumoSRON is developing X-ray transition edge sensor (TES) calorimeters arrays, as a backup technology for X-IFU instrument on the ATHENA space observatory. These detectors are based on a superconducting TiAu bilayer TES with critical temperature of 100 mK on a 1 μm thick SiN membrane with Au or Au/Bi absorbers. Number of devices have been fabricated and measured using a Frequency Division Multiplexing (FDM) readout system with 1-5 MHz bias frequencies. We measured IV curves, critical temperature, thermal conductance, noise and also X-ray energy resolution at number of selected bias points. So far our best calorimeter shows 3.9 eV X-ray resolution at 6 keV. Here we present a summary of our results and the latest status of development of X-ray calorimeters at SRON.
Referência(s)