Development of TiAu TES x-ray calorimeters for the X-IFU on ATHENA space observatory

2018; Linguagem: Inglês

10.1117/12.2313439

Autores

P. Khosropanah, E. Taralli, L. Gottardi, C. P. de Vries, K. Nagayoshi, M. Ridder, H. Akamatsu, M. P. Bruijn, J. R. Gao,

Tópico(s)

Superconducting Materials and Applications

Resumo

SRON is developing X-ray transition edge sensor (TES) calorimeters arrays, as a backup technology for X-IFU instrument on the ATHENA space observatory. These detectors are based on a superconducting TiAu bilayer TES with critical temperature of 100 mK on a 1 μm thick SiN membrane with Au or Au/Bi absorbers. Number of devices have been fabricated and measured using a Frequency Division Multiplexing (FDM) readout system with 1-5 MHz bias frequencies. We measured IV curves, critical temperature, thermal conductance, noise and also X-ray energy resolution at number of selected bias points. So far our best calorimeter shows 3.9 eV X-ray resolution at 6 keV. Here we present a summary of our results and the latest status of development of X-ray calorimeters at SRON.

Referência(s)