Artigo Acesso aberto Revisado por pares

Implementation of the cobas Liat influenza point-of-care test into an emergency department during a high-incidence season: a retrospective evaluation following real-world implementation

2018; Elsevier BV; Volume: 101; Issue: 3 Linguagem: Inglês

10.1016/j.jhin.2018.12.008

ISSN

1532-2939

Autores

Jonathan Youngs, Yasir Iqbal, Stephen Glass, Peter Riley, Cassie F. Pope, Tim Planche, D Carrington,

Tópico(s)

Pneumonia and Respiratory Infections

Resumo

The cobas Liat influenza A/B and respiratory syncytial virus (RSV) assay (Liat) was used in the adult emergency department of a large London hospital from 21st January 2018 to 14th April 2018. Influenza was detected in 308 of 1027 (30%) samples tested; influenza A in 157 (15.3%), influenza B in 149 (14.5%) and RSV in 28 (2.7%). When compared against Fast Track Diagnostics Respiratory Pathogens 21 multiplex polymerase chain reaction and Cepheid Xpert Xpress Flu/RSV assay, Liat performance for the detection of influenza A or B was: sensitivity 85% [95% confidence interval (CI) 76-92)], specificity 98% (95% CI 97-99), negative predictive value 94% (95% CI 92-96) and positive predictive value 95% (95% CI 91-97).

Referência(s)