Resumo

Abstract Micromegas detector is considered to be a promising candidate for a large variety of high-rate experiments. Micromegas of various geometries have already been established as appropriate for these experiments for their performances in terms of gas gain uniformity, energy and space point resolution, and their capability to efficiently pave large read-out surfaces with minimum dead zone. The present work investigates the effect of spacers on different detector characteristics of Micromegas detectors having various amplification gaps and mesh hole pitches. Numerical simulation has been used as a tool of exploration to evaluate the effect of such dielectric material on detector performance. Some of the important and fundamental characteristics such as electron transparency, gain and signal of the Micromegas detector have been estimated.