Artigo Revisado por pares

Paper-Based Microfluidic Device (DON-Chip) for Rapid and Low-Cost Deoxynivalenol Quantification in Food, Feed, and Feed Ingredients

2019; American Chemical Society; Volume: 4; Issue: 11 Linguagem: Inglês

10.1021/acssensors.9b01895

ISSN

2379-3694

Autores

Qian Jiang, Jiandong Wu, Yao Kang, Yulong Yin, Max M. Gong, Chengbo Yang, Francis Lin,

Tópico(s)

Advanced Chemical Sensor Technologies

Resumo

Mycotoxin contamination causes over $5 billion of economic loss per year in the North American food and feed industry. A rapid, low-cost, portable, and reliable method for on-site detection of deoxynivalenol (DON), a representative mycotoxin predominantly occurring in grains, would be helpful to control mycotoxin contamination. In this study, a paper-based microfluidic chip capable of measuring DON (DON-Chip) in food, feed, and feed ingredients was developed. The DON-Chip incorporated a colorimetric competitive immunoassay into a paper microfluidic device and used gold nanoparticles as a signal indicator. Furthermore, a novel ratiometric analysis method was proposed to improve detection resolvability. Detection of DON in the aqueous extracts from solid food, feed, or feed ingredients was successfully validated with a detection range of 0.01–20 ppm (using dilution factors from 10 to 104). Compared with conventional methods, the DON-Chip method could greatly reduce the cost and time of mycotoxin detection in the food and feed industry.

Referência(s)
Altmetric
PlumX