Long-time clinical experience in patient setup for several particle therapy clinical indications: management of patient positioning and evaluation of setup reproducibility and stability

2019; Wiley; Volume: 93; Issue: 1107 Linguagem: Inglês

10.1259/bjr.20190595

ISSN

1748-880X

Autores

R. Ricotti, Andrea Pella, B. Tagaste, Giovanni Elisei, Giulia Fontana, Maria Bonora, Mario Ciocca, Francesca Valvo, Roberto Orecchia, G. Baroni,

Tópico(s)

Radiation Dose and Imaging

Resumo

Accurate patient positioning is crucial in particle therapy due to the geometrical selectivity of particles. We report and discuss the National Center for Oncological Hadrontherapy (CNAO) experience in positioning accuracy and stability achieved with solid thermoplastic masks fixed on index base plates and assessed by daily orthogonal X-ray imaging.Positioning data were retrospectively collected (between 2012 and 2018) and grouped according to the treated anatomical site. 19696 fractions of 1325 patients were evaluated.The study was designed to assess:(i) the number of fractions in which a single correction vector was applied(SCV);(ii) the number of fractions in which further setup verification was performed (SV);(iii) the number of fractions in which SV lead to an additional correction within (MCV 5min) 5 minutes from the first setup correction;(iv) the systematic (Σ) and random (σ) error components of the correction vectors applied.A SCV was applied in 71.5% of fractions, otherwise SV was required. In 30.6% of fractions with SV, patient position was not further revised. In the remaining fractions, MCV 5min were applied mainly in extracranial and cranial sites respectively.Interfraction Σ was ≤ 1.7 mm/0.7° and σ was ≤ 1.2 mm/0.6° in cranial sites while in extracranial sites Σ was ≤ 5.5 mm/0.9° and σ was ≤4.4 mm/0.9°. Setup residuals were submillimetric in all sites. In cranial patients, maximum intrafractional Σ was 0.8 mm/0.4°.This report extensively quantifies inter- and intrafraction setup accuracy on an institutional basis and confirms the need of image guidance to fully benefit from the geometrical selectivity of particles.The reported analysis provides a board institutional data set on the evaluation of patient immobilization and bony anatomy alignment for several particle therapy clinical indications.

Referência(s)