Improved Sensitivity in the Measurement of Residual Leukocytes in Platelet Products Using an Automated Leukocyte Counter
2000; Wiley; Volume: 79; Issue: 1 Linguagem: Inglês
10.1046/j.1423-0410.2000.7910034.x
ISSN1423-0410
AutoresH. Nagahasi, Katsumi Yabusaki, Hideto Matsui, Taisuke Hirono, N. Yokokawa, Masahiro Satake, Kenji Tadokoro, Takeo Juji,
Tópico(s)Blood groups and transfusion
ResumoAbstract Background and Objectives : Leukocytes remaining in platelet concentrates may be responsible for side effects caused by platelet transfusion. A simple method of high sensitivity for determining trace numbers of leukocytes is currently needed. Materials and Methods : An automated leukocyte counter, the LD‐1000, the principle of which is a combination of spining down on the observation field of stained nuclei derived from a 100 μl neat sample and image digitization with a charge coupled device camera, was newly developed and tested for sensitivity and reproducibility. Results : While the theoretical lower limit of detection was 0.01 cells/μl, the limit of detection of the LD‐1000 was verified to be 0.2 cells/μl. A good correlation (r = 0.86) was observed between the results obtained using the Nageotte method and the LD‐1000. Repeated measurements also confirmed satisfactory reproducibility. Conclusions : The instrument provides a new method of enumerating residual leukocytes in platelet products with better sensitivity and easier procedure compared to the Nageotte method and will be useful for quality assurance purposes at blood centers.