Artigo Acesso aberto Revisado por pares

Systemic IgG Exposure and Safety in Patients with Primary Immunodeficiency: A Randomized Crossover Study Comparing a Novel Investigational Wearable Infusor Versus the Crono Pump

2022; Future Medicine; Volume: 14; Issue: 16 Linguagem: Inglês

10.2217/imt-2022-0097

ISSN

1750-7448

Autores

Richard L. Wasserman, Charlotte Cunningham‐Rundles, John Anderson, Patricia L. Lugar, Michael Palumbo, Niraj Patel, Jutta Hofmann, Fiona Glassman, Eileen Rogers, Michaela Praus, Mikhail Rojavin,

Tópico(s)

Pharmaceutical studies and practices

Resumo

Aim: A novel, Investigational Wearable Infusor (IWI) was evaluated in a randomized, controlled, crossover, open-label study to determine if its delivery of subcutaneous immunoglobulin (IgPro20) achieved a comparable area under the concentration-time curve (AUC) for immunoglobulin G (IgG) versus the Crono S-PID-50 infusion pump (CP). EudraCT: 2016-003798-16. Materials & methods: Patients with primary immunodeficiency (PID) were randomized to receive IgPro20 in Sequence 1 (CP/IWI) or 2 (IWI/CP). The primary end point was AUC for IgG during the final week of each 4-week period. Results: 23 patients were enrolled. Evaluation of area under the concentration-time curve from time 0 (pre-infusion) to 7 days after infusion (AUC0-7 days) (IWI: 1806 h*g/l; CP: 1829 h*g/l) and geometric mean ratio indicated comparable AUCs for IgG for both devices. Conclusion: Similar IgG exposure, indicated by AUC values, can be achieved with IgPro20 using the IWI or CP in PID.Patients with primary immunodeficiency (PID) are at a higher risk of developing serious infections than healthy individuals. Immunoglobulin G (IgG) replacement therapy reduces this risk, as it raises a patient's antibody levels to help fight off infections. IgG replacement therapy can be performed as an intravenous or subcutaneous (under the skin) infusion. The subcutaneous route is associated with improved quality of life for patients, as therapy can be carried out at home by the patient, allowing for more flexibility, convenience and autonomy. Wearable drug-delivery systems are devices that stick to the body and automatically deliver doses of a drug to the patient. In this study, we investigated whether a novel Investigational Wearable Infusor device could deliver the subcutaneous IgG replacement therapy, IgPro20, in patients with PID. We show that the new infusion device can deliver IgG replacement therapy and allows for similar levels of IgG to be achieved in patients as a comparator device. This wearable drug delivery device simplifies drug administration and could help address some of the challenges associated with self-injection such as complicated infusion preparation, needle phobia and concerns about pain. Trial Registration Number: 2016-003798-16 (EudraCT).

Referência(s)
Altmetric
PlumX