Subcutaneous-implantable cardioverter defibrillator lead dislodgement in a juvenile catecholamine-induced polymorphic ventricular tachycardia patient
2022; Oxford University Press; Volume: 2022; Issue: 12 Linguagem: Inglês
10.1093/omcr/omac130
ISSN2053-8855
AutoresMiwa Miyoshi, Hajime Saeki, Yo Arita, Yoshinori Iida, Tomoki Fukui, Shohei Yamamoto, Kana Shichijo, Yuto Suetani, Kosuke Hirose, M Kuramoto, Nobuyuki Ogasawara,
Tópico(s)Cardiac electrophysiology and arrhythmias
ResumoCatecholaminergic polymorphic ventricular tachycardia (CPVT) is a relatively rare inherited arrhythmic disease that causes sudden cardiac death, and is caused by mutations in the cardiac ryanodine receptor (RyR2) or sarcoplasmic reticulum protein calsequestrin 2 gene (CASQ2). A 16-year-old man was diagnosed with CPVT and was implanted with a Subcutaneous-implantable Cardioverter Defibrillator (S-ICD), but defibrillation electrode detachment occurred early after placement. We suspected that a two-incision technique was the possible cause. We also report on changes in surface ECG in remote monitoring of the device. TAKE HOME MESSAGE Although two-incision techniques are becoming the mainstream method of S-ICD implantation, we should consider that the three-incision technique may be advantageous in highly active patients. Remote monitoring may also be useful for early detection of S-ICD dislodgement.
Referência(s)