A non-systematic signal-correction error in the commercial EasyOne Pro multiple-breath washout device significantly impacts outcomes in children and adults

2022; Thieme Medical Publishers (Germany); Linguagem: Inglês

10.1055/s-0042-1754481

ISSN

1439-3824

Autores

M-A Oestreich, Florian Wyler, Blaise Etter, KA Ramsey, Philipp Latzin,

Tópico(s)

Airway Management and Intubation Techniques

Resumo

Introduction Multiple-breath nitrogen washout is an established and sensitive technique to assess functional residual capacity and ventilation inhomogeneity in the lung. Indirect calculation of nitrogen concentration requires precise calibration and accurate measurement of gas concentrations. We investigated the accuracy of the carbon dioxide gas concentration used for the indirect calculation of nitrogen concentration in a commercial multiple-breath washout (MBW) device (EasyOne Pro LAB, ndd Medizintechnik AG, Switzerland) as a possible cause of lack of comparability between outcomes of different MBW devices.